| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.252 | 2.156 | 
| 500.00 | 1.738 | 1.642 | 
| 250.00 | 1.136 | 1.040 | 
| 125.00 | 0.752 | 0.656 | 
| 62.50 | 0.593 | 0.497 | 
| 31.25 | 0.377 | 0.281 | 
| 15.63 | 0.221 | 0.125 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 94-107% | 100% | 
| EDTA plasma(n=5) | 82-95% | 88% | 
| Heparin plasma(n=5) | 94-105% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-97% | 87-96% | 82-95% | 92-105% | 
| EDTA plasma(n=5) | 85-92% | 79-96% | 89-101% | 85-97% | 
| Heparin plasma(n=5) | 86-94% | 93-102% | 89-97% | 85-94% |