| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.121 | 2.040 |
| 100.00 | 1.674 | 1.593 |
| 50.00 | 1.069 | 0.988 |
| 25.00 | 0.697 | 0.616 |
| 12.50 | 0.538 | 0.457 |
| 6.25 | 0.322 | 0.241 |
| 3.13 | 0.230 | 0.149 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-103% | 96% |
| EDTA plasma(n=5) | 78-96% | 87% |
| Heparin plasma(n=5) | 86-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-99% | 95-106% | 87-101% | 82-97% |
| EDTA plasma(n=5) | 79-92% | 80-93% | 81-98% | 89-102% |
| Heparin plasma(n=5) | 87-97% | 86-92% | 89-103% | 89-98% |