| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 3000.00 | 2.126 | 2.030 | 
| 1500.00 | 1.527 | 1.431 | 
| 750.00 | 1.035 | 0.939 | 
| 375.00 | 0.752 | 0.656 | 
| 187.50 | 0.567 | 0.471 | 
| 93.75 | 0.329 | 0.233 | 
| 46.88 | 0.208 | 0.112 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-93% | 86% | 
| EDTA plasma(n=5) | 95-108% | 101% | 
| Heparin plasma(n=5) | 82-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-93% | 91-101% | 87-98% | 95-104% | 
| EDTA plasma(n=5) | 92-101% | 82-98% | 89-97% | 85-93% | 
| Heparin plasma(n=5) | 86-101% | 86-101% | 87-96% | 92-105% |