| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 200.00 | 2.126 | 2.037 | 
| 100.00 | 1.653 | 1.564 | 
| 50.00 | 1.116 | 1.027 | 
| 25.00 | 0.878 | 0.789 | 
| 12.50 | 0.523 | 0.434 | 
| 6.25 | 0.361 | 0.272 | 
| 3.13 | 0.183 | 0.094 | 
| 0.00 | 0.089 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 95-103% | 99% | 
| EDTA plasma(n=5) | 92-102% | 97% | 
| Heparin plasma(n=5) | 81-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-96% | 85-97% | 89-101% | 85-97% | 
| EDTA plasma(n=5) | 79-93% | 89-97% | 85-94% | 87-101% | 
| Heparin plasma(n=5) | 86-94% | 85-95% | 89-99% | 85-95% |