 
| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.013 | 1.922 | 
| 500.00 | 1.542 | 1.451 | 
| 250.00 | 1.136 | 1.045 | 
| 125.00 | 0.792 | 0.701 | 
| 62.50 | 0.593 | 0.502 | 
| 31.25 | 0.377 | 0.286 | 
| 15.63 | 0.261 | 0.170 | 
| 0.00 | 0.091 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 93% | 
| EDTA plasma(n=5) | 78-92% | 85% | 
| Heparin plasma(n=5) | 92-107% | 99% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-102% | 83-96% | 97-105% | 85-97% | 
| EDTA plasma(n=5) | 97-105% | 86-92% | 95-103% | 92-104% | 
| Heparin plasma(n=5) | 85-94% | 92-101% | 87-96% | 82-90% |