| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.054 | 1.968 | 
| 5.00 | 1.624 | 1.538 | 
| 2.50 | 1.135 | 1.049 | 
| 1.25 | 0.752 | 0.666 | 
| 0.63 | 0.457 | 0.371 | 
| 0.32 | 0.403 | 0.317 | 
| 0.16 | 0.256 | 0.170 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 93% | 
| EDTA plasma(n=5) | 87-95% | 91% | 
| Heparin plasma(n=5) | 93-107% | 100% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 78-102% | 83-96% | 79-93% | 88-97% | 
| EDTA plasma(n=5) | 88-97% | 90-99% | 87-103% | 82-97% | 
| Heparin plasma(n=5) | 85-94% | 87-103% | 82-94% | 78-90% |