| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.237 | 2.134 |
| 100.00 | 1.670 | 1.567 |
| 50.00 | 1.129 | 1.026 |
| 25.00 | 0.940 | 0.837 |
| 12.50 | 0.552 | 0.449 |
| 6.25 | 0.346 | 0.243 |
| 3.13 | 0.231 | 0.128 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-92% | 88% |
| EDTA plasma(n=5) | 94-103% | 99% |
| Heparin plasma(n=5) | 81-115% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-92% | 80-104% | 85-101% | 90-111% |
| EDTA plasma(n=5) | 94-99% | 85-96% | 82-96% | 94-118% |
| Heparin plasma(n=5) | 87-98% | 85-109% | 86-115% | 80-106% |