| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 500.00 | 2.388 | 2.295 | 
| 250.00 | 1.658 | 1.565 | 
| 125.00 | 1.106 | 1.013 | 
| 62.50 | 0.835 | 0.742 | 
| 31.25 | 0.563 | 0.470 | 
| 15.63 | 0.385 | 0.292 | 
| 7.82 | 0.235 | 0.142 | 
| 0.00 | 0.093 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 90-105% | 97% | 
| EDTA plasma(n=5) | 86-99% | 92% | 
| Heparin plasma(n=5) | 79-93% | 86% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-102% | 85-96% | 90-98% | 82-93% | 
| EDTA plasma(n=5) | 89-103% | 79-92% | 88-97% | 85-97% | 
| Heparin plasma(n=5) | 95-102% | 97-105% | 87-94% | 93-101% |