Products

Human F5(Coagulation Factor V) ELISA Kit
One-click to copy product information
ELK1765
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Human F5(Coagulation Factor V) ELISA Kit
Reactivity: Human
Alternative Names: FV; FVL; PCCF; Proaccelerin; Labile Factor; Factor V Leiden; Accelerator Globulin; Activated protein C cofactor; Proaccelerin, labile factor
Assay Type: Sandwich
Sensitivity: 59 pg/mL
Standard: 8000 pg/mL
Detection Range: 125-8000 pg/mL
Sample type: plasma
Assay length: 3.5h
Research Area: Enzyme & Kinase;Metabolic pathway;Hematology;
Uniprot ID: P12259
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human F5. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human F5. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human F5, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human F5 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
8000.00 2.043 1.962
4000.00 1.527 1.446
2000.00 1.255 1.174
1000.00 0.958 0.877
500.00 0.655 0.574
250.00 0.361 0.280
125.00 0.230 0.149
0.00 0.081 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant F5 and the recovery rates were calculated by comparing the measured value to the expected amount of F5 in samples.
Matrix Recovery range Average
serum(n=5) 95-107% 101%
EDTA plasma(n=5) 77-95% 86%
Heparin plasma(n=5) 84-97% 90%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of F5 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-98% 83-96% 95-104% 81-95%
EDTA plasma(n=5) 87-98% 92-104% 82-96% 95-102%
Heparin plasma(n=5) 87-98% 92-105% 83-92% 89-98%