| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.054 | 1.972 |
| 500.00 | 1.624 | 1.542 |
| 250.00 | 1.102 | 1.020 |
| 125.00 | 0.837 | 0.755 |
| 62.50 | 0.474 | 0.392 |
| 31.25 | 0.335 | 0.253 |
| 15.63 | 0.198 | 0.116 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-93% | 86% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 86-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-104% | 83-96% | 95-103% | 88-95% |
| EDTA plasma(n=5) | 86-95% | 93-103% | 87-98% | 85-96% |
| Heparin plasma(n=5) | 92-102% | 81-93% | 86-98% | 91-103% |