| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.961 | 1.873 |
| 500.00 | 1.537 | 1.449 |
| 250.00 | 1.074 | 0.986 |
| 125.00 | 0.785 | 0.697 |
| 62.50 | 0.517 | 0.429 |
| 31.25 | 0.388 | 0.300 |
| 15.63 | 0.259 | 0.171 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-119% | 106% |
| EDTA plasma(n=5) | 94-115% | 105% |
| Heparin plasma(n=5) | 81-98% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-119% | 92-110% | 91-107% | 83-116% |
| EDTA plasma(n=5) | 95-105% | 94-101% | 81-93% | 92-103% |
| Heparin plasma(n=5) | 93-112% | 91-105% | 95-100% | 91-114% |