| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.201 | 2.100 |
| 500.00 | 1.593 | 1.492 |
| 250.00 | 1.121 | 1.020 |
| 125.00 | 0.892 | 0.791 |
| 62.50 | 0.547 | 0.446 |
| 31.25 | 0.324 | 0.223 |
| 15.63 | 0.129 | 0.028 |
| 0.00 | 0.101 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 77-95% | 86% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |
| EDTA plasma(n=5) | 86-100% | 93-108% | 80-101% | 87-96% |
| Heparin plasma(n=5) | 96-105% | 95-103% | 95-103% | 87-98% |