| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 300.00 | 2.132 | 2.049 |
| 150.00 | 1.593 | 1.510 |
| 75.00 | 0.935 | 0.852 |
| 37.50 | 0.752 | 0.669 |
| 18.75 | 0.543 | 0.460 |
| 9.38 | 0.357 | 0.274 |
| 4.69 | 0.257 | 0.174 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 86-99% | 92% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-104% | 82-96% | 92-105% | 83-92% |
| EDTA plasma(n=5) | 87-93% | 85-95% | 89-99% | 85-95% |
| Heparin plasma(n=5) | 89-97% | 85-94% | 79-96% | 85-99% |