| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.258 | 2.181 |
| 500.00 | 1.653 | 1.576 |
| 250.00 | 1.147 | 1.070 |
| 125.00 | 0.895 | 0.818 |
| 62.50 | 0.513 | 0.436 |
| 31.25 | 0.356 | 0.279 |
| 15.63 | 0.189 | 0.112 |
| 0.00 | 0.077 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 88-101% | 94% |
| Heparin plasma(n=5) | 91-105% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 85-97% | 93-102% | 81-95% |
| EDTA plasma(n=5) | 83-98% | 92-103% | 86-97% | 93-101% |
| Heparin plasma(n=5) | 87-96% | 86-98% | 91-103% | 87-90% |