| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.145 | 2.040 |
| 1000.00 | 1.491 | 1.386 |
| 500.00 | 1.120 | 1.015 |
| 250.00 | 0.830 | 0.725 |
| 125.00 | 0.530 | 0.425 |
| 62.50 | 0.361 | 0.256 |
| 31.25 | 0.215 | 0.110 |
| 0.00 | 0.105 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-116% | 104% |
| EDTA plasma(n=5) | 84-120% | 102% |
| Heparin plasma(n=5) | 92-116% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-115% | 89-102% | 91-103% | 87-105% |
| EDTA plasma(n=5) | 82-116% | 83-94% | 94-106% | 86-104% |
| Heparin plasma(n=5) | 89-95% | 91-112% | 88-116% | 86-100% |