| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 10000.00 | 0.189 | |
| 5000.00 | 0.386 | |
| 2500.00 | 0.575 | |
| 1250.00 | 0.823 | |
| 625.00 | 1.112 | |
| 312.50 | 1.537 | |
| 156.25 | 1.775 | |
| 0.00 | 2.226 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-99% | 91% |
| EDTA plasma(n=5) | 79-91% | 85% |
| Heparin plasma(n=5) | 81-97% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-95% | 85-98% | 82-97% | 92-101% |
| EDTA plasma(n=5) | 83-92% | 91-96% | 85-93% | 95-101% |
| Heparin plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |