| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 2.059 | 1.970 | 
| 2500.00 | 1.536 | 1.447 | 
| 1250.00 | 1.107 | 1.018 | 
| 625.00 | 0.821 | 0.732 | 
| 312.50 | 0.535 | 0.446 | 
| 156.25 | 0.315 | 0.226 | 
| 78.13 | 0.206 | 0.117 | 
| 0.00 | 0.089 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-95% | 89% | 
| EDTA plasma(n=5) | 86-99% | 92% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-92% | 83-96% | 93-102% | 87-98% | 
| EDTA plasma(n=5) | 85-97% | 90-99% | 82-101% | 93-105% | 
| Heparin plasma(n=5) | 95-104% | 82-96% | 92-103% | 90-99% |