| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.919 | 1.818 |
| 2500.00 | 1.542 | 1.441 |
| 1250.00 | 1.285 | 1.184 |
| 625.00 | 0.958 | 0.857 |
| 312.50 | 0.602 | 0.501 |
| 156.25 | 0.324 | 0.223 |
| 78.13 | 0.211 | 0.110 |
| 0.00 | 0.101 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 94-105% | 98% |
| Heparin plasma(n=5) | 94-107% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-106% | 93-105% | 85-99% | 90-99% |
| EDTA plasma(n=5) | 95-106% | 87-101% | 82-97% | 87-102% |
| Heparin plasma(n=5) | 81-94% | 93-107% | 93-102% | 91-105% |