| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 100.00 | 1.809 | 1.707 | 
| 50.00 | 1.558 | 1.456 | 
| 25.00 | 1.183 | 1.081 | 
| 12.50 | 0.827 | 0.725 | 
| 6.25 | 0.406 | 0.304 | 
| 3.13 | 0.377 | 0.275 | 
| 1.57 | 0.216 | 0.114 | 
| 0.00 | 0.102 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-97% | 90% | 
| EDTA plasma(n=5) | 86-99% | 92% | 
| Heparin plasma(n=5) | 80-92% | 86% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 93-102% | 81-96% | 93-106% | 82-95% | 
| EDTA plasma(n=5) | 88-103% | 86-97% | 87-102% | 89-101% | 
| Heparin plasma(n=5) | 87-98% | 92-101% | 82-98% | 89-97% |