| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.161 | 2.070 |
| 5.00 | 1.557 | 1.466 |
| 2.50 | 1.151 | 1.060 |
| 1.25 | 0.874 | 0.783 |
| 0.63 | 0.536 | 0.445 |
| 0.31 | 0.318 | 0.227 |
| 0.16 | 0.171 | 0.080 |
| 0.00 | 0.091 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-108% | 99% |
| EDTA plasma(n=5) | 81-93% | 87% |
| Heparin plasma(n=5) | 89-102% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-99% | 92-117% | 80-88% | 92-118% |
| EDTA plasma(n=5) | 81-110% | 87-110% | 83-98% | 91-97% |
| Heparin plasma(n=5) | 88-105% | 86-93% | 94-115% | 82-100% |