| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.231 | 2.145 | 
| 500.00 | 1.697 | 1.611 | 
| 250.00 | 1.235 | 1.149 | 
| 125.00 | 0.811 | 0.725 | 
| 62.50 | 0.551 | 0.465 | 
| 31.25 | 0.393 | 0.307 | 
| 15.63 | 0.237 | 0.151 | 
| 0.00 | 0.086 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 86-99% | 92% | 
| EDTA plasma(n=5) | 82-95% | 88% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 79-91% | 95-104% | 93-102% | 79-94% | 
| EDTA plasma(n=5) | 87-98% | 92-104% | 82-96% | 95-102% | 
| Heparin plasma(n=5) | 90-99% | 83-97% | 81-103% | 89-101% |