| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.898 | 1.812 |
| 5.00 | 1.517 | 1.431 |
| 2.50 | 1.043 | 0.957 |
| 1.25 | 0.828 | 0.742 |
| 0.63 | 0.456 | 0.370 |
| 0.32 | 0.322 | 0.236 |
| 0.16 | 0.165 | 0.079 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-96% | 88% |
| EDTA plasma(n=5) | 82-98% | 90% |
| Heparin plasma(n=5) | 84-98% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-99% | 83-101% | 88-96% | 90-101% |
| EDTA plasma(n=5) | 85-101% | 84-98% | 85-93% | 86-98% |
| Heparin plasma(n=5) | 86-93% | 91-105% | 87-98% | 89-94% |