| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.152 | 2.070 |
| 500.00 | 1.677 | 1.595 |
| 250.00 | 1.174 | 1.092 |
| 125.00 | 0.823 | 0.741 |
| 62.50 | 0.544 | 0.462 |
| 31.25 | 0.308 | 0.226 |
| 15.63 | 0.259 | 0.177 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-109% | 101% |
| EDTA plasma(n=5) | 80-88% | 84% |
| Heparin plasma(n=5) | 81-90% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-115% | 90-113% | 91-97% | 91-108% |
| EDTA plasma(n=5) | 86-120% | 81-93% | 87-104% | 91-119% |
| Heparin plasma(n=5) | 81-90% | 94-115% | 90-106% | 81-94% |