| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.136 | 2.068 | 
| 500.00 | 1.629 | 1.561 | 
| 250.00 | 1.115 | 1.047 | 
| 125.00 | 0.757 | 0.689 | 
| 62.50 | 0.421 | 0.353 | 
| 31.25 | 0.257 | 0.189 | 
| 15.63 | 0.174 | 0.106 | 
| 0.00 | 0.068 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 85-103% | 94% | 
| EDTA plasma(n=5) | 95-105% | 100% | 
| Heparin plasma(n=5) | 93-101% | 97% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-97% | 85-94% | 87-96% | 85-97% | 
| EDTA plasma(n=5) | 85-94% | 87-96% | 92-101% | 82-98% | 
| Heparin plasma(n=5) | 87-91% | 85-93% | 89-97% | 85-93% |