| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.226 | 2.143 |
| 500.00 | 1.629 | 1.546 |
| 250.00 | 1.118 | 1.035 |
| 125.00 | 0.885 | 0.802 |
| 62.50 | 0.440 | 0.357 |
| 31.25 | 0.290 | 0.207 |
| 15.63 | 0.162 | 0.079 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-111% | 101% |
| EDTA plasma(n=5) | 86-111% | 99% |
| Heparin plasma(n=5) | 90-105% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-89% | 91-110% | 93-106% | 85-118% |
| EDTA plasma(n=5) | 83-97% | 94-115% | 88-95% | 89-96% |
| Heparin plasma(n=5) | 94-117% | 90-106% | 83-104% | 90-112% |