| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.204 | 2.119 |
| 500.00 | 1.675 | 1.590 |
| 250.00 | 1.217 | 1.132 |
| 125.00 | 0.897 | 0.812 |
| 62.50 | 0.595 | 0.510 |
| 31.25 | 0.335 | 0.250 |
| 15.63 | 0.235 | 0.150 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-87% | 83% |
| EDTA plasma(n=5) | 80-91% | 86% |
| Heparin plasma(n=5) | 96-105% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 78-84% | 87-94% | 96-105% | 79-92% |
| EDTA plasma(n=5) | 84-96% | 78-87% | 90-96% | 90-96% |
| Heparin plasma(n=5) | 91-100% | 80-86% | 90-104% | 89-96% |