| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.142 | 2.046 |
| 1000.00 | 1.681 | 1.585 |
| 500.00 | 1.244 | 1.148 |
| 250.00 | 0.852 | 0.756 |
| 125.00 | 0.606 | 0.510 |
| 62.50 | 0.352 | 0.256 |
| 31.25 | 0.185 | 0.089 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-115% | 98% |
| EDTA plasma(n=5) | 86-124% | 105% |
| Heparin plasma(n=5) | 94-120% | 107% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-107% | 93-113% | 89-110% | 85-109% |
| EDTA plasma(n=5) | 88-120% | 95-109% | 91-99% | 86-96% |
| Heparin plasma(n=5) | 91-108% | 91-98% | 90-118% | 81-110% |