| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.219 | 2.132 | 
| 5.00 | 1.643 | 1.556 | 
| 2.50 | 1.154 | 1.067 | 
| 1.25 | 0.896 | 0.809 | 
| 0.63 | 0.498 | 0.411 | 
| 0.32 | 0.364 | 0.277 | 
| 0.16 | 0.206 | 0.119 | 
| 0.00 | 0.087 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 87-99% | 93% | 
| EDTA plasma(n=5) | 93-107% | 100% | 
| Heparin plasma(n=5) | 87-95% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-97% | 86-92% | 95-103% | 98-106% | 
| EDTA plasma(n=5) | 95-102% | 87-101% | 82-91% | 85-97% | 
| Heparin plasma(n=5) | 93-102% | 78-92% | 87-98% | 80-93% |