| 48 T | $370.00 | 
| 96 T | $528.00 | 
| 96 T*5 | $2244.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.368 | 2.260 | 
| 500.00 | 1.674 | 1.566 | 
| 250.00 | 1.119 | 1.011 | 
| 125.00 | 0.854 | 0.746 | 
| 62.50 | 0.496 | 0.388 | 
| 31.25 | 0.321 | 0.213 | 
| 15.63 | 0.225 | 0.117 | 
| 0.00 | 0.108 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-95% | 88% | 
| EDTA plasma(n=5) | 94-107% | 100% | 
| Heparin plasma(n=5) | 94-105% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 79-91% | 96-105% | 80-92% | 91-103% | 
| EDTA plasma(n=5) | 85-94% | 87-96% | 82-95% | 81-93% | 
| Heparin plasma(n=5) | 85-92% | 79-96% | 95-102% | 87-98% |