| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.154 | 2.052 |
| 5.00 | 1.532 | 1.430 |
| 2.50 | 1.208 | 1.106 |
| 1.25 | 0.882 | 0.780 |
| 0.63 | 0.536 | 0.434 |
| 0.31 | 0.341 | 0.239 |
| 0.16 | 0.243 | 0.141 |
| 0.00 | 0.102 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-100% | 94% |
| EDTA plasma(n=5) | 82-120% | 101% |
| Heparin plasma(n=5) | 95-114% | 105% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-101% | 94-118% | 93-105% | 93-119% |
| EDTA plasma(n=5) | 90-108% | 83-98% | 84-115% | 86-100% |
| Heparin plasma(n=5) | 92-108% | 80-120% | 83-97% | 84-106% |