| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.214 | 2.125 |
| 1000.00 | 1.782 | 1.693 |
| 500.00 | 1.152 | 1.063 |
| 250.00 | 0.847 | 0.758 |
| 125.00 | 0.563 | 0.474 |
| 62.50 | 0.395 | 0.306 |
| 31.25 | 0.251 | 0.162 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 94-107% | 100% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-105% | 86-102% | 97-106% | 89-97% |
| EDTA plasma(n=5) | 78-94% | 87-96% | 85-93% | 88-103% |
| Heparin plasma(n=5) | 85-92% | 90-96% | 87-1028% | 85-97% |