| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.186 | 2.088 |
| 250.00 | 1.598 | 1.500 |
| 125.00 | 1.120 | 1.022 |
| 62.50 | 0.817 | 0.719 |
| 31.25 | 0.524 | 0.426 |
| 15.63 | 0.391 | 0.293 |
| 7.81 | 0.253 | 0.155 |
| 0.00 | 0.098 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-114% | 99% |
| EDTA plasma(n=5) | 80-111% | 96% |
| Heparin plasma(n=5) | 82-103% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-100% | 93-117% | 85-107% | 90-113% |
| EDTA plasma(n=5) | 87-102% | 85-93% | 91-114% | 90-106% |
| Heparin plasma(n=5) | 85-119% | 82-97% | 84-118% | 81-118% |