| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.940 | 1.832 |
| 5.00 | 1.602 | 1.494 |
| 2.50 | 1.302 | 1.194 |
| 1.25 | 0.933 | 0.825 |
| 0.63 | 0.584 | 0.476 |
| 0.31 | 0.383 | 0.275 |
| 0.16 | 0.225 | 0.117 |
| 0.00 | 0.108 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-113% | 101% |
| EDTA plasma(n=5) | 80-99% | 90% |
| Heparin plasma(n=5) | 87-120% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-102% | 91-105% | 86-93% | 90-116% |
| EDTA plasma(n=5) | 86-105% | 94-107% | 84-113% | 95-118% |
| Heparin plasma(n=5) | 84-115% | 82-103% | 92-101% | 86-107% |