| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 30.00 | 2.153 | 2.061 |
| 15.00 | 1.751 | 1.659 |
| 7.50 | 1.290 | 1.198 |
| 3.75 | 0.935 | 0.843 |
| 1.88 | 0.589 | 0.497 |
| 0.94 | 0.405 | 0.313 |
| 0.47 | 0.243 | 0.151 |
| 0.00 | 0.092 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-117% | 103% |
| EDTA plasma(n=5) | 92-126% | 109% |
| Heparin plasma(n=5) | 88-112% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-100% | 82-105% | 92-107% | 82-96% |
| EDTA plasma(n=5) | 91-99% | 83-94% | 88-119% | 93-109% |
| Heparin plasma(n=5) | 92-116% | 93-109% | 83-111% | 86-112% |