48 T | $265.00 |
96 T | $376.00 |
Concentration (U/mL) | OD | Corrected OD |
---|---|---|
20.00 | 2.168 | 2.079 |
10.00 | 1.529 | 1.440 |
5.00 | 0.973 | 0.884 |
2.50 | 0.618 | 0.529 |
1.25 | 0.451 | 0.362 |
0.63 | 0.285 | 0.196 |
0.32 | 0.158 | 0.069 |
0.00 | 0.089 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 95-108% | 101% |
EDTA plasma(n=5) | 82-97% | 89% |
Heparin plasma(n=5) | 84-93% | 87% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-96% | 86-95% | 90-102% | 81-90% |
EDTA plasma(n=5) | 87-95% | 83-96% | 92-102% | 82-93% |
Heparin plasma(n=5) | 83-96% | 81-94% | 86-98% | 92-101% |