| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.128 | 2.041 | 
| 500.00 | 1.709 | 1.622 | 
| 250.00 | 1.116 | 1.029 | 
| 125.00 | 0.943 | 0.856 | 
| 62.50 | 0.506 | 0.419 | 
| 31.25 | 0.351 | 0.264 | 
| 15.63 | 0.225 | 0.138 | 
| 0.00 | 0.087 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-98% | 90% | 
| EDTA plasma(n=5) | 86-99% | 92% | 
| Heparin plasma(n=5) | 80-95% | 87% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-99% | 83-95% | 98-104% | 96-106% | 
| EDTA plasma(n=5) | 85-100% | 84-98% | 95-101% | 88-102% | 
| Heparin plasma(n=5) | 86-93% | 91-103% | 87-98% | 95-101% |