| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.953 | 1.862 |
| 500.00 | 1.523 | 1.432 |
| 250.00 | 1.136 | 1.045 |
| 125.00 | 0.845 | 0.754 |
| 62.50 | 0.524 | 0.433 |
| 31.25 | 0.345 | 0.254 |
| 15.63 | 0.199 | 0.108 |
| 0.00 | 0.091 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-93% | 88% |
| EDTA plasma(n=5) | 84-106% | 95% |
| Heparin plasma(n=5) | 88-94% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-114% | 94-120% | 88-96% | 94-101% |
| EDTA plasma(n=5) | 90-119% | 94-117% | 88-96% | 91-119% |
| Heparin plasma(n=5) | 92-102% | 85-102% | 85-116% | 91-107% |