| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.043 | 1.947 |
| 5.00 | 1.514 | 1.418 |
| 2.50 | 1.129 | 1.033 |
| 1.25 | 0.869 | 0.773 |
| 0.63 | 0.591 | 0.495 |
| 0.31 | 0.393 | 0.297 |
| 0.16 | 0.225 | 0.129 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-107% | 96% |
| EDTA plasma(n=5) | 87-122% | 105% |
| Heparin plasma(n=5) | 82-102% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-112% | 94-117% | 83-114% | 95-118% |
| EDTA plasma(n=5) | 82-100% | 88-102% | 88-117% | 83-104% |
| Heparin plasma(n=5) | 82-95% | 86-102% | 90-99% | 88-109% |