| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.236 | 2.151 |
| 50.00 | 1.771 | 1.686 |
| 25.00 | 1.147 | 1.062 |
| 12.50 | 0.834 | 0.749 |
| 6.25 | 0.555 | 0.470 |
| 3.13 | 0.364 | 0.279 |
| 1.57 | 0.189 | 0.104 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-94% | 91% |
| EDTA plasma(n=5) | 92-106% | 99% |
| Heparin plasma(n=5) | 78-87% | 83% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-90% | 90-103% | 81-94% | 78-91% |
| EDTA plasma(n=5) | 92-103% | 96-102% | 82-91% | 82-91% |
| Heparin plasma(n=5) | 94-107% | 82-95% | 89-102% | 96-105% |