| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.035 | 1.940 |
| 500.00 | 1.717 | 1.622 |
| 250.00 | 1.194 | 1.099 |
| 125.00 | 0.831 | 0.736 |
| 62.50 | 0.582 | 0.487 |
| 31.25 | 0.363 | 0.268 |
| 15.63 | 0.258 | 0.163 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-105% | 98% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-97% | 86-95% | 81-93% | 91-103% |
| EDTA plasma(n=5) | 85-99% | 82-96% | 92-101% | 82-90% |
| Heparin plasma(n=5) | 79-92% | 80-93% | 81-98% | 89-102% |