| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.179 | 2.097 |
| 500.00 | 1.576 | 1.494 |
| 250.00 | 1.224 | 1.142 |
| 125.00 | 0.876 | 0.794 |
| 62.50 | 0.585 | 0.503 |
| 31.25 | 0.324 | 0.242 |
| 15.63 | 0.258 | 0.176 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 77-85% | 81% |
| EDTA plasma(n=5) | 90-97% | 94% |
| Heparin plasma(n=5) | 87-93% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 96-106% | 89-102% | 84-94% | 88-95% |
| EDTA plasma(n=5) | 81-95% | 84-90% | 82-93% | 82-93% |
| Heparin plasma(n=5) | 92-101% | 83-95% | 96-106% | 91-102% |