| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.154 | 2.067 |
| 5.00 | 1.790 | 1.703 |
| 2.50 | 1.174 | 1.087 |
| 1.25 | 0.752 | 0.665 |
| 0.63 | 0.541 | 0.454 |
| 0.32 | 0.325 | 0.238 |
| 0.16 | 0.190 | 0.103 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-93% | 86% |
| EDTA plasma(n=5) | 80-97% | 88% |
| Heparin plasma(n=5) | 96-109% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-96% | 91-107% | 87-101% | 95-102% |
| EDTA plasma(n=5) | 92-103% | 83-98% | 88-97% | 96-104% |
| Heparin plasma(n=5) | 89-101% | 84-101% | 95-104% | 88-105% |