| 48 T | $265.00 | 
| 96 T | $376.00 | 
| 96 T*5 | $1598.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.075 | 1.992 | 
| 500.00 | 1.636 | 1.553 | 
| 250.00 | 1.186 | 1.103 | 
| 125.00 | 0.847 | 0.764 | 
| 62.50 | 0.517 | 0.434 | 
| 31.25 | 0.319 | 0.236 | 
| 15.63 | 0.129 | 0.046 | 
| 0.00 | 0.083 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 84-98% | 91% | 
| EDTA plasma(n=5) | 89-103% | 96% | 
| Heparin plasma(n=5) | 82-95% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 86-95% | 91-106% | 87-100% | 95-101% | 
| EDTA plasma(n=5) | 95-102% | 83-97% | 88-96% | 96-103% | 
| Heparin plasma(n=5) | 85-97% | 84-100% | 95-103% | 88-104% |