| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 8000.00 | 2.074 | 1.959 |
| 4000.00 | 1.622 | 1.507 |
| 2000.00 | 1.273 | 1.158 |
| 1000.00 | 0.968 | 0.853 |
| 500.00 | 0.542 | 0.427 |
| 250.00 | 0.356 | 0.241 |
| 125.00 | 0.217 | 0.102 |
| 0.00 | 0.115 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-95% | 91% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 92-106% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-93% | 91-104% | 87-98% | 95-99% |
| EDTA plasma(n=5) | 89-99% | 83-95% | 88-94% | 96-101% |
| Heparin plasma(n=5) | 85-100% | 84-98% | 95-101% | 88-102% |