| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.027 | 1.940 |
| 50.00 | 1.685 | 1.598 |
| 25.00 | 1.385 | 1.298 |
| 12.50 | 0.828 | 0.741 |
| 6.25 | 0.508 | 0.421 |
| 3.13 | 0.298 | 0.211 |
| 1.57 | 0.190 | 0.103 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-106% | 100% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 87-93% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 93-107% | 93-102% | 91-105% |
| EDTA plasma(n=5) | 92-106% | 93-105% | 85-99% | 90-99% |
| Heparin plasma(n=5) | 95-106% | 87-101% | 82-97% | 87-102% |