| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.179 | 2.083 |
| 5.00 | 1.527 | 1.431 |
| 2.50 | 1.207 | 1.111 |
| 1.25 | 0.936 | 0.840 |
| 0.63 | 0.794 | 0.698 |
| 0.32 | 0.362 | 0.266 |
| 0.16 | 0.174 | 0.078 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-95% | 91% |
| EDTA plasma(n=5) | 78-90% | 84% |
| Heparin plasma(n=5) | 84-97% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-97% | 93-101% | 97-103% | 91-99% |
| EDTA plasma(n=5) | 81-97% | 88-102% | 85-93% | 83-97% |
| Heparin plasma(n=5) | 87-102% | 87-96% | 92-101% | 83-96% |