| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.998 | 1.909 |
| 5.00 | 1.639 | 1.550 |
| 2.50 | 1.093 | 1.004 |
| 1.25 | 0.855 | 0.766 |
| 0.63 | 0.479 | 0.390 |
| 0.32 | 0.356 | 0.267 |
| 0.16 | 0.173 | 0.084 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 82-98% | 90% |
| Heparin plasma(n=5) | 90-107% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-106% | 85-101% | 87-102% | 89-95% |
| EDTA plasma(n=5) | 88-97% | 86-92% | 91-99% | 79-97% |
| Heparin plasma(n=5) | 85-104% | 79-92% | 83-95% | 86-97% |