| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 3.188 | 3.127 |
| 500.00 | 2.626 | 2.565 |
| 250.00 | 1.737 | 1.676 |
| 125.00 | 0.965 | 0.904 |
| 62.50 | 0.454 | 0.393 |
| 31.25 | 0.231 | 0.170 |
| 15.63 | 0.157 | 0.096 |
| 0.00 | 0.061 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 96-107% | 102% |
| EDTA plasma(n=5) | 96-105% | 101% |
| Heparin plasma(n=5) | 80-92% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 78-85% | 81-95% | 88-99% | 85-91% |
| EDTA plasma(n=5) | 92-104% | 86-100% | 86-93% | 86-93% |
| Heparin plasma(n=5) | 86-93% | 82-92% | 92-103% | 84-93% |