| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 300.00 | 3.167 | 3.107 |
| 150.00 | 1.849 | 1.789 |
| 75.00 | 1.264 | 1.204 |
| 37.50 | 0.713 | 0.653 |
| 18.75 | 0.387 | 0.327 |
| 9.38 | 0.217 | 0.157 |
| 4.69 | 0.157 | 0.097 |
| 0.00 | 0.060 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 77-85% | 81% |
| EDTA plasma(n=5) | 85-92% | 89% |
| Heparin plasma(n=5) | 93-104% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-92% | 88-99% | 79-88% | 88-101% |
| EDTA plasma(n=5) | 80-86% | 87-97% | 78-84% | 78-84% |
| Heparin plasma(n=5) | 82-95% | 79-88% | 84-91% | 92-100% |