| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 3000.00 | 2.487 | 2.432 |
| 1500.00 | 1.643 | 1.588 |
| 750.00 | 1.188 | 1.133 |
| 375.00 | 0.728 | 0.673 |
| 187.50 | 0.394 | 0.339 |
| 93.75 | 0.243 | 0.188 |
| 46.88 | 0.136 | 0.081 |
| 0.00 | 0.055 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-104% | 97% |
| EDTA plasma(n=5) | 92-103% | 98% |
| Heparin plasma(n=5) | 82-89% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-95% | 90-98% | 83-96% | 96-104% |
| EDTA plasma(n=5) | 84-95% | 87-93% | 90-96% | 90-96% |
| Heparin plasma(n=5) | 89-96% | 77-87% | 78-87% | 95-108% |